To be attached long hour observation equipment using your grown cell.
Setting temp up at 37°C, and “temp check up” in central chamber part is executed gradually.
“Temp in chamber” is controlled within ±0.5°C of set up.
Comparison chart to normal incubation on the cell increase
| PHOENIX | CO2 Incubator | |||
|---|---|---|---|---|
| Incubated hour | Cell density(cells/mL) | Useful cell rate(%) | Cell density(cell/mL) | Useful cell rate(%) |
| 0hour | 9.37×105 | 93% | 9.84×105 | 93% |
| 18hours | 1.67×106 | 93% | 1.72×106 | 93% |
| 48hours | 3.22×106 | 93% | 3.25×106 | 95% |
| n=6 | ||||
Setting up inside environment of phoenix chamber at 37°C and 5% CO2, cell density and useful cell rate of Jurkat cell are measured gradually.